Filter material dynamic filtration performance tester experimental process

The filter material dynamic filtration performance tester is mainly used to test the filtration performance of filter materials. The equipment itself is easy to operate, compact, and sturdy in design. The dust concentration value of the filtration test can be adjusted quickly and conveniently, and the actual use simulation degree is high, simulating a variety of actual conditions. Afterward, the particle size and concentration are measured online by the digital system of the light scattering spectrometer.


As a modern technology enterprise integrating R&D, manufacturing, sales, training, and service, Shanghai Qianshi Precision Electromechanical Technology Co., Ltd is committed to delivering more testing instruments to the market, providing testing instruments for textiles, leather, combustion, automotive interior and exterior trims, material environmental climate aging, etc., and non-standard customization is also acceptable. Targeted tests can be carried out according to customer requirements before purchase to ensure that the purchased instruments are suitable.

Comply with the standards:
GB/T6719-2009 Technical requirements for bag dust collectors
ISO 11057-2011 Filter performance test of cleanable filter media
VDI 3926-2004 Filter performance test of cleanable filter media

Main features:
1. The dynamic performance detection device for filter materials has the advantages of reasonable structural design, accurate test parameters, and simple and easy operation. It can complete comprehensive performance tests such as resistance and efficiency of filter material cleanliness, aging, and stability.
2. Color LCD operation interface, digital graphics processing, and XDT software can independently display the Chinese interface and edit and process the curve. It has two functions: custom and programmable, and comes with 5 standard test methods. It has a fault self-calibration function. It has an operation encryption function and can be remotely controlled.

Experimental process:
1. Initialization treatment: Determination of dust filtration performance of initial filter material sample, cleaning is performed when pressure loss reaches 1000Pa, and the weight gain of high-efficiency filter membrane and outlet dust concentration is measured and recorded after 30 repetitions;
2. Aging treatment: 10,000 aging treatments with 5-second intervals;
3. Stabilization treatment: Stabilization treatment, installation process A performs 10 dust filtration and cleaning operations;
4. Stabilization treatment: Perform 30 dust filtration and cleaning operations according to A.



Experimental parameters:

Particle type: non-sticky dry powder

Continuous operation time: 1-50 hours
Compressed air requirements: maximum 6bar
Dust mass flow: 0.05g/h-20g/h, continuously adjustable, stability less than 5%
Aerosol outlet mass concentration: 0.012g-13g/h
Power requirements: 100-260VAC/47-63Hz
Filter material test area: 140mm
Filter wind speed: 1-3m/min adjustable
Horizontal pipe flow: 0-3m³/h
Extensible bypass pipe
Pressure sensor: 0-2000Pa, accuracy 2%
Backflush pressure: 0.2-0.6MPa (and can be set by yourself)
Backflush time: 50ms-200ms (and can be set by yourself)
With a time-controlled pulse valve and pressure-controlled pulse valve
Suction flow controller: mass flow meter 80LPM, accuracy 0.1LPM
Dust dispersion compressed air control: Mass flow meter 20LPM, accuracy 0.1LPM
Vacuum pump: 8m³/h, 4X10-4mbar, 550W
Filter material filtration performance control software: capable of sampling pressure signals for A, B, C, and D filtration processes; automatic control of constant pressure injection; injection control during aging process; with manual, semi-automatic, and automatic testing
Control mode: NEU data processing mode.