Automated Filter Tester


The Automated Filter Tester Model G506 continues to be the best solution for testing particulate respirator filters, disposable filtering face pieces, and a wide assortment of filter media. The filter testers have been used in quality control and manufacturing by leading filter and filter media manufacturers and test centers around the world for more than 20 years.
 
Based on established model G506, the world's most-used filter tester, model G506  now offers higher sensitivity and resolution - up to 99.9999 percent efficiency has been produced with Emery oil at concentrations of 200 mg/m? - at a lower cost of ownership due to the new serviceable photometers inside. The model G506 features a high degree of automation and self-diagnostics that greatly simplifies operation, increases throughput, and improves overall measurement performance. Whether looking for a stand-alone tester to check a few filters per day or need to integrate a tester in an automated, high-volume production line, the model G506 is up to the challenge.
 
The built-in capability to test with salt and oil means that just one unit is needed to test your product to:
US 42 CFR 84, JMOL, ISO 16900-3, EN 143    ISO 23328-1 (and more),NIOSH 42 CFR Part 84,JMOL,GB 19083 ,GB2626,GB 32610,GB19082,YY0469,YY0969.
 
Features and Benefits:
Complies to more standards than any other filter tester
Salt and oil testing in one unit 
Highly consistent test results
New user-interface and data recording options
Customizable fully automated testing sequences
Ergonomic design
Reduced cost of ownership with new serviceable photometers
Worldwide service and support
 
Applications:
USA commercial respirator regulation (42 CFR part 84)
European EN 143, ISO 16900-3 and related respirator standards 
View EN 143 Compliance Report (US)
View EN 143 Compliance Report (DE)
ISO 23328-1 (testing filters for anaesthetic use)
JMOL (Japanese respirator standard)
And, many more national respirator regulations
Filter media testing
Consult your TSI representative for more details
 
Specifications:
Items Parameter
Apllication Load and quality control
Max Efficiency 99.999%
Aerosol Type DOP/PAO/DEHS/Paraffi/Others Oil Aerosol、Salt
Aerosol Generation Atomizing
CMD 0.2μm(Oil)or 0.75um(NaCl)
GSD <1.6(Oil)or <1.86(NaCl)
Flow Rate 15-100L/min
Resistance 0-150mm H2O(0-1470Pa)
Particle Detect Scatter Photometer
Measurement Time 10 Sec.
Data Report LCD,RS-232
Operation Independent tewster、automatic production line
Standard compliant US 42 CFR part 84,EN 143,JMOL,ISO 23328-1 ISO 16900-3
Power Supply 115VAC、60Hz,10A或230VAC,50Hz,5A
Compressor Air 80psig,7scfm(550kPa,198std.L/min)
Dimension 70cm x 70cm x 147cm(28in, x 28in.x 58in.)